商品描述
MODEL HCC-215 · High power · 240 VAC · 15 AMP · 3600 watts
The line of HCC controllers is designed to power large scale equipment with volumes up to 100 L.
Built with the same uncompromising commitment to safety as our Model 270, the HCC is available in three models for both 120 and 230 volt operations. In addition to the features of the Model 270, these high power controllers incorporate a 100 hour timer to turn heating OFF (or ON) at a user set time. The HCC is readily customized to meet unique research or additional safety needs. Call for more information.
Includes a built-in, independent, backup controller to guard against heating accidents from equipment failure. The desired temperature is entered into the main temperature controller which regulates heating, and then a high temperature cut-off is entered into the backup limit controller. If the reaction reaches the high temperature limit or a thermocouple breaks, power is turned off to the heater until the controller is manually reset. Both meters independently monitor the reaction temperature. In the event that one meter fails, the other takes over to prevent a heating accident. J-KEM’s design provides independent, 100% redundant control.
Contains J-KEM’s patented power control computer which provides 0.1º C regulation of anything.
Requires one dual element thermocouple and cord, or two single element.
Product Description
1. Power Regulation
J-KEM’s original power control computer is replaced by a next generation microprocessor. The power control computer is J-KEM’s patented technology that adjusts power to the heater 2048 times per second resulting in 0.1o C regulation.
2. USB Communications
PC communications and free KEM-Net software enables remote PC control, GLP/GMP compliant data collection, and multi-temperature ramps built in an Excel-like table.
3. Upgradable Programming
Software modules can be uploaded to the processor, at no charge, adding new capabilities to the controller. For example, automatic detection of exothermic reactions.
J-KEM’s patented power control computer makes this the only controller that can heat volumes under 100 µL with the safety of 0.1°C regulation.
Additional Information
Features |
|
Specifications | 208 - 240vac, 15 amps, 3600 watts Dimensions 5.5” x 12” x 12.5.” (H x W x D) |